User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: Validyne MC1-3:Three Channel Module Case DATE: 24-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 28 NUMBER OF LINES: 223 CONFIGURATION: HP 34401A STANDARD: Clarostat 240-C STANDARD: DTB Metered Variac ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 STD Clarostat 240-C 1.004 STD DTB Metered Variac 1.005 IEEE [@34401]*RST[GTL] 1.006 HEAD INITIAL CONDITIONS 1.007 DISP The test adaptor consists of two cards. 1.007 DISP One labled VOLTMETER and the othe labled LOAD. 1.008 DISP When directed to make connections for the DVM, these 1.008 DISP connections MUST be made on the card labled VOLTMETER. 1.009 DISP When directed to make connections for the Clarostat, 1.009 DISP these connections must be made on the card labled LOAD. 1.010 DISP Plug the test adaptor in to one of the UUT slots. 1.011 DISP Plug the UUT power cord into to the variac output. 1.012 DISP Turn on the variac and set its output for 115 volts . 1.013 DISP Turn on the UUT. # 1.012 JMPT 14.003 Validyne MC1-20:Twenty Channel Module Cas 1.014 HEAD {} 1.015 HEAD {* DC Power Supply Tests *} 1.016 HEAD {* Load Regulation *} 1.017 JMP 2.003 1.018 EVAL 2.001 DISP Remove the Clarostat from the test adaptor. 2.002 JMP 2.004 2.003 DISP Connect the 34401 to the GRD. and +15 terminals. 2.004 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 2.005 MEME 2.006 DISP Set the Clarostat to 100 ohm and connect it to the 2.006 DISP GRD and +15 terminal posts. 2.007 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 2.008 MEME 2.009 MEMC V 0.100% 3.001 JMP 4.003 3.002 EVAL 4.001 DISP Remove the Clarostat from the -15 terminal. 4.002 JMP 4.004 4.003 DISP Move the 33401 HI connection to the -15 terminal post. 4.004 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 4.005 MEME 4.006 DISP Move the Clarostat hi connection to the -15 terminal. 4.007 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 4.008 MEME 4.009 MEMC V 0.100% 5.001 HEAD {} 5.002 HEAD {* Line Regulation *} 5.003 JMP 6.001 5.004 EVAL # 2.002 DISP Connect the 34401 to the +15 and GRD. terminals # 2.002 DISP on the test adaptor. 6.001 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 6.002 MEME >1 6.003 DISP Set the variac for 104 volts. 6.004 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 6.005 MEME 6.006 MEME <1 6.007 MEMC V 0.100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 7.001 DISP Set the variac for 127 volts. 7.002 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 7.003 MEME 7.004 MEME <1 7.005 MEMC V 0.100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 8.001 DISP Set the variac for 115 volts. 8.002 DISP Move the 33401 HI connection to the +15 terminal post. 8.003 DISP Move the Clarostat hi connection to the +15 terminal. 8.004 JMP 9.002 8.005 EVAL 9.001 DISP Reset the variac for 115 volts 9.002 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 9.003 MEME >1 9.004 DISP Set the variac for 104 volts. 9.005 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 9.006 MEME 9.007 MEME <1 9.008 MEMC V 0.100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 10.001 JMP 11.001 10.002 EVAL 11.001 DISP Set the variac for 127 volts. 11.002 IEEE [@34401]MEAS:VOLT:DC?[I][GTL] 11.003 MEME 11.004 MEME <1 11.005 MEMC V 0.100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 12.001 DISP Set the variac for 115 volts. 12.002 DISP Disconnect the Clarostat from the UUT. 12.003 HEAD {} 12.004 HEAD {* Carrier Tests *} 12.005 JMP 13.001 12.006 EVAL 13.001 DISP Connect the 34401 to the AC terminals on the adaptor. 13.001 DISP Make all connections to the card inside the UUT. 13.002 IEEE [@34401]CONF:VOLT:AC 13.003 ACC 5.000V 0.01% 13.004 IEEE [@34401]MEAS:VOLT:AC?[T0][I][GTL] 13.005 MEME 13.006 MEMC V 0.100U #! Test Tol 0.1, Sys Tol 0.0005, TUR 200.000 (>= 4.00). 14.001 IEEE [@34401]MEAS:FREQ?[T0][I][GTL] 14.002 MEM/ 1000 14.003 MEME >255 14.004 ACC 3.000kHz 10P% 14.005 MEME 14.006 MEMC kHz 1.000% #! Test Tol 0.025, Sys Tol 3e-005, TUR 833.333 (>= 4.00). 15.001 HEAD {} 15.002 HEAD {* Load Regulation Test *} 15.003 JMP 16.002 15.004 EVAL 16.001 DISP Remove the Clarostat from the AC terminal. 16.002 IEEE [@34401]MEAS:VOLT:AC?[D2000][I][GTL] 16.003 MEME 16.004 JMPT 16.007 Validyne MC1-20:Twenty Channel Module Cas 16.005 DISP Set the Clarostat to 100 ohms and connect from one 16.005 DISP AC terminal to GRD. Make this connection on the LOAD 16.005 DISP board. 16.006 JMP 16.008 16.007 DISP Set the Clarostat to 1 ohm and connect from one 16.007 DISP AC terminal to GRD. Make this connection on the LOAD 16.007 DISP board. 16.008 IEEE [@34401]MEAS:VOLT:AC?[D2000][I][GTL] 16.009 JMPT 17.002 Validyne MC1-20:Twenty Channel Module Cas 16.010 MEMC V 0.100% 17.001 JMP 18.001 17.002 MEMC V 0.500% #! T.U.R. not calculated because System Uncertainty not available. 18.001 HEAD {} 18.002 HEAD {* Line Regulation Test *} 18.003 JMP 19.001 18.004 EVAL 19.001 DISP Make sure the variac is set at 115 volts. 19.002 IEEE [@34401]MEAS:VOLT:AC?[D2000][I][GTL] 19.003 MEME >1 19.004 DISP Set the variac to 104 volts. 19.005 IEEE [@34401]MEAS:VOLT:AC?[D2000][I][GTL] 19.006 MEME 19.007 MEME <1 19.008 MEMC V 0.0100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 20.001 DISP Set the variac to 127 volts. 20.002 IEEE [@34401]MEAS:VOLT:AC?[D2000][I][GTL] 20.003 MEME 20.004 MEME <1 20.005 MEMC V 0.0100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 21.001 DISP Set the variac to 115 volts. 21.002 HEAD {} 21.003 HEAD {* Frequency Load Regulation Test *} 21.004 JMP 22.001 21.005 EVAL # 22.001 IEEE [@34401]MEAS:FREQ?[T0][I][GTL] # 22.002 MEM/ 1000 22.001 MEME <255 22.002 MEME #20.004 DISP Set the Clarostat to 100 ohms and connect from one #20.004 DISP AC terminal to GRD. 22.003 IEEE [@34401]MEAS:FREQ?[T0][I][GTL] 22.004 MEM/ 1000 22.005 MEMC kH 0.100% #! T.U.R. not calculated because System Uncertainty not available. 23.001 HEAD {} 23.002 HEAD {* Frequency Line Regulation Test *} 23.003 JMP 24.001 23.004 EVAL 24.001 DISP Make sure the variac is set at 115 volts. 24.002 IEEE [@34401]MEAS:FREQ?[T0][I][GTL] 24.003 MEM/ 1000 24.004 MEME >1 24.005 DISP Set the variac to 104 volts. 24.006 IEEE [@34401]MEAS:FREQ?[T0][I][GTL] 24.007 MEM/ 1000 24.008 MEME 24.009 MEME <1 24.010 MEMC kH 0.0100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 25.001 DISP Set the variac to 127 volts. 25.002 IEEE [@34401]MEAS:FREQ?[T0][I][GTL] 25.003 MEM/ 1000 25.004 MEME 25.005 MEME <1 25.006 MEMC kH 0.0100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 26.001 DISP Set the variac to 115 volts. 26.002 DISP Disconnect the 34401A from the test adaptor. 26.003 HEAD {} 26.004 HEAD {* Distortion Test *} 26.005 JMP 27.001 26.006 EVAL 27.001 DISP Connect the 8903A INPUT HI BNC to the AC terminals on 27.001 DISP the adaptor. 27.002 IEEE [@8903][TERM LF] 27.003 IEEE [@8903]L1 27.004 IEEE [@8903]M3 27.005 IEEE [@8903]LN 27.006 IEEE [@8903][I] 27.007 ACC 0.00pct 0.01U 27.008 MEME 27.009 MEMC pct 1U 28.001 DISP Turn off the variac. 28.002 DISP Disconnect all test equipment from the UUT. 28.003 IEEE [@34401]*RST[GTL] 28.004 HEAD {} 28.005 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 12 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.